Removal of large foreign bodies made of thermoplastic material from the eye

Abstract

Purpose

To present the results of research aimed at developing a new method of minimally invasive removal of an intraocular foreign body (IOFB) from the eye cavity.

Methods

The main idea was to fix the IOFB onto the edge of a laser probe, which is transsclerally injected into the eye. After that, the IOFB is fixed at the end of the laser probe and removed transpupillary from the eye through an incision in the cornea. In order to fix the IOFB on the laser probe edge, this edge should be brought closely to the IOFB surface. The material of the IOFB should be locally fused using the laser emission. During this process, the probe edge should be submerged into the liquid melt. After laser emission is turned off, the material around the probe is hardened and the IOFB is hardly fixed on the laser probe edge.

Results and Conclusions

The main laser emission parameters and their influence on the eye tissue were investigated within a thermodynamic model. It is shown that for IOFB made from polyethylene terephthalate the optimal laser power is close to 2 W, the rate of rise of light energy is more than 10 ms, and the wavelength is approximately 0.905 μm. With these parameters (taking into account multiple scattering effects), the light impact on the retina will not exceed 100 mW/cm2, which is not dangerous for eye tissue. Efficiency is supported by clinical testing. The method was used for intraocular extraction of an Airsoft ball bearing (6 mm sphere of polyethylene terephthalate).The overall extraction time was approximately 4 minutes.

Eur J Ophthalmol 2017; 27(5): 621 - 625

Article Type: SURGICAL TECHNIQUE

DOI:10.5301/ejo.5000945

Authors

Georgiy Stolyarenko, Dzhasser Doroshenko, Victor Salakhutdinov, Yuri Smetanin

Article History

Disclosures

Financial support: No financial support was received for this submission.
Conflict of interest: None of the authors has conflict of interest with this submission.

This article is available as full text PDF.

  • If you are a Subscriber, please log in now.

  • Article price: Eur 36,00
  • You will be granted access to the article for 72 hours and you will be able to download any format (PDF or ePUB). The article will be available in your login area under "My PayPerView". You will need to register a new account (unless you already own an account with this journal), and you will be guided through our online shop. Online purchases are paid by Credit Card through PayPal.
  • If you are not a Subscriber you may:
  • Subscribe to this journal
  • Unlimited access to all our archives, 24 hour a day, every day of the week.

Authors

Affiliations

  • Posterior Eye Segment Diagnostics and Surgery Center, Moscow - Russia
  • Scientific Research Institute for System Studies, Russian Academy of Sciences, Moscow - Russia
  • Federal Research Center “Informatics and Control”, Russian Academy of Sciences, Moscow - Russia

Article usage statistics

The blue line displays unique views in the time frame indicated.
The yellow line displays unique downloads.
Views and downloads are counted only once per session.

No supplementary material is available for this article.